CLASS II ONGOING Device recall · Reported 11 Nov 2026 · FDA Enforcement Report
Veritas Reverse Total Shoulder Replacement recalled by restor3d
restor3d is recalling Veritas Reverse Total Shoulder Replacement, distributed nationwide in the US. The recall was initiated on 1 Oct 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0348-2027
- FDA event ID
- 100047
- Recalling firm
- restor3d, Durham, NC
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 1 Oct 2026
- FDA classified
- 3 Nov 2026
- Reported
- 11 Nov 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 1 Oct 2026 | Recall initiated by company | |
| 3 Nov 2026 | Classified by FDA | +33 days |
| 11 Nov 2026 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
Product as listed by the FDA
Veritas Reverse Total Shoulder Replacement, Model/Catalog Number: 3500-140C3, 3500-140U3, 3500-140C1, 3500-140U1, size 1 -2, 40mm Polymer Liner
Where it was distributed
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
Nationwide ColoradoFloridaMassachusettsNorth CarolinaOklahomaTexas
Questions about this recall
Why was it recalled?
(Retroactively reported) Pairing of shoulder implant stem sizes 1 and 2 with the largest glenosphere diameter (40 mm) creates an excessive mechanical loading configuration in which certain polyethylene components lack adequate retention strength, potentially resulting in dislocation, reduced range of motion, or revision surgery.
Which lots are affected?
All Lot codes/ UDIs:Product Code: 3500-140C3 UDI/GTIN:00840097513492; Product Code: 3500-140U3 UDI/GTIN:00840097513461; Product Code: 3500-140C1 UDI/GTIN:00840097513430; Product Code: 3500-140U1 UDI/GTIN:00840097513409
Where was it sold?
US Nationwide distribution in the states of Texas, Massachusetts, Oklahoma, Florida, Colorado, North Carolina.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 17 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from restor3d
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Metatarsophalangeal (MTP) Implant is a sterile recalled by restor3d | restor3d | Sterility | Nationwide |
| Device | CLASS II | Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide) recalled | restor3d | Packaging Defects | Nationwide |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Evans Wedge Disposable Trials, Large Footprint Thickness recalled | restor3d | Other | Nationwide |
| Device | CLASS II | Non-Sterile Double Ended Trial Kit, Small Footprint recalled by restor3d | restor3d | Other | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |