CLASS III Device recall · Reported 11 Nov 2015 · FDA Enforcement Report

Vitros Chemistry Products Calibrator Kit 9 Calibrator, Product recalled

Ortho-Clinical Diagnostics is recalling Vitros Chemistry Products Calibrator Kit 9 Calibrator, Product, distributed nationwide in the US. The recall was initiated on 28 Sep 2015 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesLot 0954, exp. date 31-May-2016
Quantity recalledDomestic: 2443 kits; Foreign: 1548 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0202-2016
FDA event ID
72289
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class III
Status
Terminated
Recall initiated
28 Sep 2015
FDA classified
3 Nov 2015
Reported
11 Nov 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

28 Sep 2015Recall initiated by company
3 Nov 2015Classified by FDA+36 days
11 Nov 2015Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm is conducting a product correction due to an increase in U90-382 or 6LU condition codes (wash errors) when using VITROS Calibrator Kit 9, Lot 0954.

Product as listed by the FDA

VITROS Chemistry Products Calibrator Kit 9 Calibrator, 2 mL, Product Code 8568040, Unique Device Identifier No. 10758750005031, IVD -- Ortho Clinical Diagnostics, Inc. --- VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System for the quantitative measurement of ACET, CRBM, DGXN, and PHYT.

Where it was distributed

Nationwide, Puerto Rico and foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela

Nationwide Puerto Rico

Questions about this recall

Is Vitros Chemistry Products Calibrator Kit 9 Calibrator, Product recalled?

Yes. Ortho-Clinical Diagnostics recalled Vitros Chemistry Products Calibrator Kit 9 Calibrator, Product on 28 Sep 2015. The recall is Class III and its status is Terminated. Recall number Z-0202-2016.

Why was it recalled?

The firm is conducting a product correction due to an increase in U90-382 or 6LU condition codes (wash errors) when using VITROS Calibrator Kit 9, Lot 0954.

Which lots are affected?

Lot 0954, exp. date 31-May-2016

Where was it sold?

Nationwide, Puerto Rico and foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656