CLASS II ONGOING Device recall · Reported 16 Nov 2022 · FDA Enforcement Report

Vitros Chemistry Products Calibrator Kit 9 -(Gen) recalled

Ortho-Clinical Diagnostics is recalling Vitros Chemistry Products Calibrator Kit 9 -(Gen) 22 of Vitros Phyt Slides For in vitro, distributed nationwide in the US. The recall was initiated on 13 Oct 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 10758750005031 Kit Lots/Exp. Date: 0911 24-MAY-2023; 0921 04-OCT-2023; 0942 06-JUN-2024; 0980 26-OCT-2022; 0991 01-MAR-2023 VITROS Chemistry Products PHYT Slides, GEN 22, All lots Calibration Diskette Data Release Versions (DRVs) 6194 through 6200 supporting VITROS Calibrator Kit 9.
Quantity recalled17,396 Units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0220-2023
FDA event ID
91051
Recalling firm
Ortho-Clinical Diagnostics, Raritan, NJ
Classification
Class II
Status
Ongoing
Recall initiated
13 Oct 2022
FDA classified
10 Nov 2022
Reported
16 Nov 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

13 Oct 2022Recall initiated by company
10 Nov 2022Classified by FDA+28 days
16 Nov 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.

Product as listed by the FDA

VITROS Chemistry Products Calibrator Kit 9 -(GEN) 22 of VITROS PHYT Slides For in vitro diagnostic use only. Product Code: 8568040 . VITROS Chemistry Products Calibrator Kit 9 is used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of ACET, CRBM, DGXN, PHBR, and PHYT

Where it was distributed

Worldwide Distribution: US (Nationwide) and Countries (Foreign) including: Australia Bermuda DV04 Bermuda, HM 13 Bermuda, HM08 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden United Kingdom

Nationwide

Questions about this recall

Why was it recalled?

Calibration disk Data Release Version (DRV) 6194 to 6200 contain inappropriate calibration mathematics for VITROS Calibration Kit 9 for calibrating VITROS PHYT Slides, for all lots of slide generation 22 (GEN 22). The erroneous mathematics on the calibration disk prevents the VITROS 250/350 Chemistry System from successfully calibrating, leading to delayed results.

Which lots are affected?

UDI-DI: 10758750005031 Kit Lots/Exp. Date: 0911 24-MAY-2023; 0921 04-OCT-2023; 0942 06-JUN-2024; 0980 26-OCT-2022; 0991 01-MAR-2023 VITROS Chemistry Products PHYT Slides, GEN 22, All lots Calibration Diskette Data Release Versions (DRVs) 6194 through 6200 supporting VITROS Calibrator Kit 9.

Where was it sold?

Worldwide Distribution: US (Nationwide) and Countries (Foreign) including: Australia Bermuda DV04 Bermuda, HM 13 Bermuda, HM08 Brazil Canada L3R 4G5 Chile China Colombia Denmark France Germany India Italy Japan Mexico Norway Portugal Russia Singapore 609917 Spain Sweden United Kingdom

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls