CLASS II Device recall · Reported 7 May 2014 · FDA Enforcement Report

Vitros Chemistry Products FS Diluent Pack recalled

Ortho-Clinical Diagnostics is recalling Vitros Chemistry Products FS Diluent Pack 3, distributed nationwide in the US. The recall was initiated on 26 Mar 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot 01-3266 (exp. 23-SEPTEMBER-2014)
Quantity recalled1042 distributed (220 went to warehouse in France)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1520-2014
FDA event ID
67883
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Terminated
Recall initiated
26 Mar 2014
FDA classified
28 Apr 2014
Reported
7 May 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

26 Mar 2014Recall initiated by company
28 Apr 2014Classified by FDA+33 days
7 May 2014Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.

Product as listed by the FDA

VITROS Chemistry Products FS Diluent Pack 3, REF 680 1754, 3 x 45 mL, (D1)/15 mL (D2), IVD --- Ortho-Clinical Diagnostics, Inc. Rochester, NY, 14626

Where it was distributed

Nationwide, Puerto Rico, and foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

Nationwide Puerto Rico

Questions about this recall

Is Vitros Chemistry Products FS Diluent Pack 3 recalled?

Yes. Ortho-Clinical Diagnostics recalled Vitros Chemistry Products FS Diluent Pack 3 on 26 Mar 2014. The recall is Class II and its status is Terminated. Recall number Z-1520-2014.

Why was it recalled?

Ortho Clinical Diagnostics (OCD) confirmed the potential for lower than expected C-Reactive Protein (CRP) results only to be generated when using VITROS Chemistry Products Specialty Diluent (Lot F3168) or VITROS Chemistry Products FS Diluent Pack 3 (Lot 01-3266). The VITROS Instructions for Use states that if the if a C-reactive protein concentration exceeds the systems measuring (reportable or dynamic) range, dilute the sample with VITROS Specialty Diluent or a patient sample containing a low concentration of CRP. An initial threefold dilution is recommended. Internal testing determined that results using VITROS CRP Slides have the potential for approximately 51% bias using a 3x dilution factor.

Which lots are affected?

Lot 01-3266 (exp. 23-SEPTEMBER-2014)

Where was it sold?

Nationwide, Puerto Rico, and foreign distribution to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.