CLASS II Device recall · Reported 10 Jul 2013 · FDA Enforcement Report

Vitros 5600 Integrated System (product recalled

Ortho-Clinical Diagnostics is recalling Vitros 5600 Integrated System (product, distributed nationwide in the US. The recall was initiated on 6 May 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSerial numbers: J56000110, J56000118 and J56001095 through J56001507
Quantity recalledTotal 381 units ( 227 units domestically & 154 units internationally)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1649-2013
FDA event ID
65207
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Terminated
Recall initiated
6 May 2013
FDA classified
1 Jul 2013
Reported
10 Jul 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

6 May 2013Recall initiated by company
1 Jul 2013Classified by FDA+56 days
10 Jul 2013Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.

Product as listed by the FDA

VITROS 5600 Integrated System (product code 6802413) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.

Where it was distributed

Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouri

Questions about this recall

Is Vitros 5600 Integrated System (product recalled?

Yes. Ortho-Clinical Diagnostics recalled Vitros 5600 Integrated System (product on 6 May 2013. The recall is Class II and its status is Terminated. Recall number Z-1649-2013.

Why was it recalled?

There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.

Which lots are affected?

Serial numbers: J56000110, J56000118 and J56001095 through J56001507

Where was it sold?

Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ortho-Clinical Diagnostics

All 311

Other Test Kits & Diagnostics recalls

All 2,656