CLASS II Device recall · Reported 10 Jul 2013 · FDA Enforcement Report
Vitros 5600 Integrated System (product recalled
Ortho-Clinical Diagnostics is recalling Vitros 5600 Integrated System (product, distributed nationwide in the US. The recall was initiated on 6 May 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1649-2013
- FDA event ID
- 65207
- Recalling firm
- Ortho-Clinical Diagnostics, Rochester, NY
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 May 2013
- FDA classified
- 1 Jul 2013
- Reported
- 10 Jul 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Test Kits & Diagnostics
Timeline
| 6 May 2013 | Recall initiated by company | |
| 1 Jul 2013 | Classified by FDA | +56 days |
| 10 Jul 2013 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Product as listed by the FDA
VITROS 5600 Integrated System (product code 6802413) Product Usage: For use in the in vitro quantitative, semi-quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Slides, VITROS Chemistry Products MicroTip Reagents and VITROS Immunodiagnostic Products Reagents.
Where it was distributed
Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouri
Questions about this recall
Is Vitros 5600 Integrated System (product recalled?
Yes. Ortho-Clinical Diagnostics recalled Vitros 5600 Integrated System (product on 6 May 2013. The recall is Class II and its status is Terminated. Recall number Z-1649-2013.
Why was it recalled?
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Which lots are affected?
Serial numbers: J56000110, J56000118 and J56001095 through J56001507
Where was it sold?
Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Ortho-Clinical Diagnostics
All 311| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Chemistry Products LAC Slides recalled by Ortho-Clinical Diagnostics | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products LAC Slides recalled | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products Ca Slides recalled | Ortho-Clinical Diagnostics | Device Malfunction | Nationwide |
| Device | CLASS II | Vitros Chemistry Products CHE Slides (Japan) recalled | Ortho-Clinical Diagnostics | Labeling Errors | Nationwide |
| Device | CLASS II | Vitros Chemistry Products CHE Slides recalled | Ortho-Clinical Diagnostics | Labeling Errors | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |