CLASS II Device recall · Reported 10 Feb 2016 · FDA Enforcement Report
ABX Pentra Glucose PAP CP recalled by Horiba Instruments
Horiba Instruments is recalling ABX Pentra Glucose PAP CP, distributed in 29 states. The recall was initiated on 11 Jan 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0740-2016
- FDA event ID
- 73033
- Recalling firm
- Horiba Instruments, Irvine, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 11 Jan 2016
- FDA classified
- 2 Feb 2016
- Reported
- 10 Feb 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 11 Jan 2016 | Recall initiated by company | |
| 2 Feb 2016 | Classified by FDA | +22 days |
| 10 Feb 2016 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Product as listed by the FDA
ABX PENTRA Glucose PAP CP ref. A11A01668 reagent is intended for the quantitative in vitro diagnostic determination of glucose in human serum, plasma and urine using glucose oxidase method by colorimetry.
Where it was distributed
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
AlabamaArkansasArizonaCaliforniaFloridaGeorgiaIdahoIllinoisKansasKentuckyMarylandMichiganMinnesotaMississippiMontanaNorth CarolinaNebraskaNew MexicoNew YorkOhio
Questions about this recall
Why was it recalled?
N-Acetylcysteine (NAC) present in the blood of patients treated for paracetamol overdose can interfere with the Trinder reaction that uses hydrogen peroxide catalytic on aminoantipyrine and phenol, and can produce falsely low results with the reagents using the Trinder reaction method.
Which lots are affected?
All Lots
Where was it sold?
U.S. distribution to the following; CA, NY, AL, OK, MN, MS, MD, TX, KY, WV, MT, IL, FL, VA, WA, MI, KS, NE, PA, OH, ID, OR, AR, AZ, NM, NC, SC, GA, TN No foreign distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
5 other products · FDA event 73033| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | ABX Pentra Triglycerides CP recalled by Horiba Instruments | Horiba Instruments | Potency & Assay | 29 states | |
| CLASS II | ABX Pentra Uric Acid CP recalled by Horiba Instruments | Horiba Instruments | Potency & Assay | 29 states | |
| CLASS II | ABX Pentra Lactic Acid CP recalled by Horiba Instruments | Horiba Instruments | Other | 29 states | |
| CLASS II | The ABX Pentra Cholesterol CP recalled by Horiba Instruments | Horiba Instruments | Potency & Assay | 29 states | |
| CLASS II | ABX Pentra Enzymatic Creatinine CP recalled by Horiba Instruments | Horiba Instruments | Other | 29 states |
More recalls from Horiba Instruments
All 28| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Horiba custom configured fluorescence instrument recalled | Horiba Instruments | Foreign Material | Nationwide |
| Device | CLASS II | Clinical Chemistry, ABX Pentra Lipase CP, ABX Pentra 400/Pentrac400 recalled | Horiba Instruments | Other | 40 states |
| Device | CLASS II | ABX Pentra Micro Albumin 2 CP, Serum/Plasma recalled | Horiba Instruments | Other | Nationwide |
| Device | CLASS II | Reagent Cup, B1037307, Sap 1221037307 for use recalled | Horiba Instruments | Device Malfunction | Nationwide |
| Device | CLASS II | Reagent Cup, B1034626, Sap 1221034626 for use recalled | Horiba Instruments | Device Malfunction | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 468| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | The MiniMed Go app for iOS or MiniMed Go app recalled | Medtronic MiniMed | Device Malfunction | Nationwide |
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG Strip recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | ANNOUNCED | StatStrip Glucose Hospital Meter System recalled by Nova BiomedicalNova Biomedical StatStrip | Nova Biomedical | Device Malfunction | |
| Device | CLASS I | StatStrip Glucose Hospital Meter System recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |