CLASS II Device recall · Reported 17 Jun 2015 · FDA Enforcement Report

Agfa Dx-d 100 is a mobile digital radiography x-ray system recalled

AGFA Healthcare is recalling Agfa Dx-d 100 is a mobile digital radiography x-ray system, distributed in 29 states. The recall was initiated on 29 May 2015 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesA5410000053, A5410000054, A5411000439, A5411000446, A5411000447, A5411000448, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5411000420, A5410000100, A5410000107, A5410000114, A5410000115, A5410000116, A5410000117, A5410000132, A5410000133, A5410000134, A5410000039, A5411000140, A5411000142, A5411000314, A5411000342, A5411000347, A5411000349, A5410000007, A5411000017, A5411000315, A5411000331, A5411000332, A5411000463, A5411000141, A5411000144, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000019, A5411000400, A5411000401, A5411000047, A5411000048, A5411000053, A5410000041, A5411000042, A5411000046, A5411000041, A5411000079, A5411000024, A5411000038, A5411000083, A5411000032, A5411000132, A5411000247, A5411000088, A5411000050, A5411000070, A5411000078, A5411000082, A5411000476, A5411000478, A5411000480, A5411000482, A5411000489, A5411000490, A5411000154, A5411000156, A5411000404, A5411000236, A5411000155, A5411000151, A5411000191, A5411000192, A5411000194, A5411000196, A5411000249, A5411000052, A5411000185, A5411000237, A5411000030, A5411000397, A5411000348, A5411000193, A5411000268, A5411000403, A5411000153, A5411000233, A5411000235, A5411000318, A5411000398, A5411000371, A5410000043, A5411000181, A5411000158, A5411000152, A5411000309, A5411000263, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000468, A5411000469, A5411000470, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000338, A5411000159, A5410000038, A5411000148, A5411000149, A5411000134, A5411000125, A5411000147, A5411000039, A5411000121, A5411000405, A5411000372, A5411000120, A5411000117, A5411000333, A5411000334, A5411000040, A5411000377, A5411000240, A5411000085, A5411000419, A5411000097, A5411000281, A5411000278, A5411000279, A5411000492, A5411000280, A5411000282, A5411000245, A5411000277, A5411000184, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000399, A5411000491, A5411000493, A5411000122, A5411000495, A5411000312, A5410000044, A5411000037, A5411000108, A5411000110, A5411000114, A5411000241, A5411000246, A5411000074, A5411000080, A5411000033, A5411000139, A5411000182, A5411000190, A5411000021, A5411000374, A5411000218, A5411000084, A5411000098, A5411000238, A5411000116, A5411000408, A5411000409, A5410000042, A5411000023, A5411000337, A5411000338, A5411000340, A5411000341, A5411000475, A5411000373, A5410000079, A5410000131, A5410000066, A5410000082, A5411000025, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000363, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000034, A5411000081, A5411000146, A5411000092, A5411000093, A5411000407, A5411000026, A5411000068, A5411000367, A5411000368, A5411000465, A5411000060, A5411000131
Quantity recalled233 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1745-2015
FDA event ID
71393
Recalling firm
AGFA Healthcare, Greenville, SC
Classification
Class II
Status
Terminated
Recall initiated
29 May 2015
FDA classified
10 Jun 2015
Reported
17 Jun 2015
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Other Reasons

Timeline

29 May 2015Recall initiated by company
10 Jun 2015Classified by FDA+12 days
17 Jun 2015Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur.

Product as listed by the FDA

AGFA DX-D 100 is a mobile digital radiography x-ray system

Where it was distributed

US Distribution to the states of : AL, CA, DC, FL, GA, IL, IA, KY, LA, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, OR, SC, TN, TX, VA, WV, WI and WY.

AlabamaCaliforniaDistrict of ColumbiaFloridaGeorgiaIowaIllinoisKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiNorth CarolinaNebraskaNew JerseyNew YorkOhio

Questions about this recall

Is Agfa Dx-d 100 is a mobile digital radiography x-ray system recalled?

Yes. AGFA Healthcare recalled Agfa Dx-d 100 is a mobile digital radiography x-ray system on 29 May 2015. The recall is Class II and its status is Terminated. Recall number Z-1745-2015.

Why was it recalled?

When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur.

Which lots are affected?

A5410000053, A5410000054, A5411000439, A5411000446, A5411000447, A5411000448, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5411000420, A5410000100, A5410000107, A5410000114, A5410000115, A5410000116, A5410000117, A5410000132, A5410000133, A5410000134, A5410000039, A5411000140, A5411000142, A5411000314, A5411000342, A5411000347, A5411000349, A5410000007, A5411000017, A5411000315, A5411000331, A5411000332, A5411000463, A5411000141, A5411000144, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000019, A5411000400, A5411000401, A5411000047, A5411000048, A5411000053, A5410000041, A5411000042, A5411000046, A5411000041, A5411000079, A5411000024, A5411000038, A5411000083, A5411000032, A5411000132, A5411000247, A5411000088, A5411000050, A5411000070, A5411000078, A5411000082, A5411000476, A5411000478, A5411000480, A5411000482, A5411000489, A5411000490, A5411000154, A5411000156, A5411000404, A5411000236, A5411000155, A5411000151, A5411000191, A5411000192, A5411000194, A5411000196, A5411000249, A5411000052, A5411000185, A5411000237, A5411000030, A5411000397, A5411000348, A5411000193, A5411000268, A5411000403, A5411000153, A5411000233, A5411000235, A5411000318, A5411000398, A5411000371, A5410000043, A5411000181, A5411000158, A5411000152, A5411000309, A5411000263, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000468, A5411000469, A5411000470, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000338, A5411000159, A5410000038, A5411000148, A5411000149, A5411000134, A5411000125, A5411000147, A5411000039, A5411000121, A5411000405, A5411000372, A5411000120, A5411000117, A5411000333, A5411000334, A5411000040, A5411000377, A5411000240, A5411000085, A5411000419, A5411000097, A5411000281, A5411000278, A5411000279, A5411000492, A5411000280, A5411000282, A5411000245, A5411000277, A5411000184, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000399, A5411000491, A5411000493, A5411000122, A5411000495, A5411000312, A5410000044, A5411000037, A5411000108, A5411000110, A5411000114, A5411000241, A5411000246, A5411000074, A5411000080, A5411000033, A5411000139, A5411000182, A5411000190, A5411000021, A5411000374, A5411000218, A5411000084, A5411000098, A5411000238, A5411000116, A5411000408, A5411000409, A5410000042, A5411000023, A5411000337, A5411000338, A5411000340, A5411000341, A5411000475, A5411000373, A5410000079, A5410000131, A5410000066, A5410000082, A5411000025, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000363, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000034, A5411000081, A5411000146, A5411000092, A5411000093, A5411000407, A5411000026, A5411000068, A5411000367, A5411000368, A5411000465, A5411000060, A5411000131

Where was it sold?

US Distribution to the states of : AL, CA, DC, FL, GA, IL, IA, KY, LA, MD, MA, MI, MN, MS, MO, NE, NJ, NY, NC, OH, OK, OR, SC, TN, TX, VA, WV, WI and WY.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from AGFA Healthcare

All 12

Other Imaging Equipment recalls

All 2,336